Reference line for accurate tray placement
Handy positioning strips hold material in place for easier wrapping
Three-layer reinforcement zones protect edges and corners
White inner sheet makes inspection up to 86% faster
Pull tabs for easy aseptic opening
Impact-resistant fabric resists tears and cuts in handling
POWERGUARD technology for 99% bacterial filtration effectiveness
Fire-resistant; abrasion-resistant
H650 suggested wrapped package content weight: 17 to 25 lbs.
SMART-FOLD* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using: Pre-vacuum steam at 270°F/132°C for 4 minutes - the wrap was validated for a dry time of 30 minutes
100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/ 55ºC and 40% – 80% relative humidity for 60 minutes
and Advanced Sterilization Products STERRAD® Sterilization System
Do not use wrap in dry heat or radiation sterilization methods